| Height |
23 inches |
| Depth |
23 inches |
| Width |
17 inches |
| Weight |
|
Free-Standing |
45 LBS |
Fully
Enclosed |
100 LBS |
| Case Material |
Military Grade Plastic |
| All other Materials |
Aluminum, brass and plastic |
| Operating Temperature Range |
35 to 110 degrees, F. |
Gas Supply Sources
02 Main and Cylinders
Air and/or Air Compressor
Oxygen Concentrator |
38-70 psi (50 psi is optimal)
38-70 psi (50 psi is optimal)
3-10 psi |
| Flowmeter Fresh Gas Flow |
1 to above 20 lpm (each
flowmeter) |
| Oxygen
Flush Valve |
Recessed,
self-closing, push button color coded and labeled,
provides 45-55 lpm contant. flow while push-button
is depressed |
| CO2 Absorber System |
King Systems KAB-9, (re-fillable)
or KAB-1 (pre-filled/disposable) |
| C02 Absorber Canister Capacity |
400 grams soda-lime |
| Directional Valves |
Built-in the CO2 absorber |
| CO2 Absorber Holding Bracket |
Plastic, secured with
knob to main frame of machine. |
| Bag-Ventilator Switch/PRV and Scavenger
Outlet Port |
Hand-operated selector
switch and rotating knob for PRV and scavenging
outlet. |
| Bellows |
Latex free, upward inflating,
range from
0 to 1.6L |
| Bellows Pressure Relief |
Pre-set at 60 cmH20 |
| Common Gas Outlet |
Size indexed, quick connect |
| Tubing Circuit |
King Systems
F-360-61 or any standard anesthesia circle circuit |
| Gas Pressure Hoses |
DISS/thread indexed, female
connectors at both ends |
| Gas Inlet Manifold |
DISS/thread indexed, male
connectors with one-way valves |
| Oxygen and Air Supply
Gauges |
0-3000 PSI reading color
coded and
clearly labeled |
| Gas Inlet Manifold
Filters |
Located behind Manifold
Air and O2 inlet male connectors. |
Gas Inlet Pressure
Regulators
Main Supply Cylinder
Safety back up Cylinder |
0-3000 PSI reading color
coded and clearly labeled
DISS/thread indexed for O2
Pin indexed, yoke mounted for "D" and
"E" cylinders for 02 |
Oxygen Supply Alarms
|
|
| Main and Safety back-up |
Pneumatically
actuated when O2 supply falls below 35 psi |
O2 Concentrator
|
Pneumatically
actuated when O2 supply falls below 1psi |
Alarm power source
|
9-volt battery
located in body of alarm box |
| Alarm on/off |
Labeled toggle
switch located on body of alarm box. |
| Air and O2 Flowmeters
|
Calibrated and scaled
0-10 LPM color coded. O2 flowmeter has a fluted
control knob for easy identification by touch |
| Oxygen Concentrator |
To power O2 flowmeter
only |
| Air Compressor
|
May be used to power ventilator
and air flowmeter |
| Oxygen Analyzer/Monitor
|
OM-25-ME (or equivalent)
Sensor life expectancy 2 years under normal conditions |
| Oxygen Analyzer Power
Source |
2 each AA batteries, life
expectancy approx. 3000 use hours |
| Auxillary
O2 Flow Selector |
Scaled
0-10 LPM in set increments, used for pre/post anesthesia |
| Vaporizer |
Penlon
SigmaDelta Series, bolt (cage) mounted, temperature
compensated, very low maintenance |
| Airway Pressure Gauge
|
Dual scaled
in cmH20 and mmHg, located on front panel of ventilator |
| Pressure Gauge Tubing |
Attached to
bag/vent switch arm |
| Mechanical
Ventilator |
Pneumatically
powered, time cycled, volume constant, pressure
variable |
| Volume Range |
0 to 1.6 L |
| Insp. Flow Range |
0 to .90 lps |
| Insp. Time Range |
0.2 to 3.0
seconds |
| Exp. Time Range |
0.2 to 30 seconds |
| Ventilator
Gas Power Requirements |
40
to 70 psi, 50 psi optimal
Use toggle switch to select gas source |
| Waste
Gas Scavenger |
Positive
and negative relief valves, 1 L reservoir bag, vacuum
control knob |
| Total
Machine Gas Leakage |
@ 30cm
H2O
@ 80cm H2O |
0ML/min
0ML/min |
| Internal
System Compliance |
@ 20cm
H2O
@ 40cm H2O |
1.1ML/cm
H2O
1.3ML/cm H2O |
| Internal
System Resistance |
@ 1.0L/sec
gas flow
@ 0.5L/sec gas flow |
4.11cm
H2O
1.80cm H2O |
| APL
Valve Pressure Drop |
@ 3.0L/min
gas flow
@ 30L/min gas flow |
0.12cm
H2O
1.03cm H2O |
| Storage |
|
| Indoor |
+ 160
°F Allow unit to warm to room temp. |
Indoor
Outdoor |
-30
°F Allow unit to warm to room temp. |