| Height |
23 inches |
| Depth |
23 inches |
| Width |
17 inches |
| Weight |
|
Free-Standing |
45 LBS |
Fully
Enclosed |
130 LBS |
| Case Material |
Military Grade Plastic |
| All other Materials |
Aluminum, brass and Plastic, military grade |
| Operating Temperature Range |
35 degrees F to 110 degrees F |
Gas Supply Sources
02 Main and Cylinders
Air and/or Air Compressor
Oxygen Concentrator |
38-70 psi (50 psi is optimal)
38-70 psi (50 psi is optimal)
3 to 10 psi (For O2 Flowmeter only) |
| Flowmeter Fresh Gas Flow |
1 to above 20 lpm (each
flowmeter) |
| Oxygen
Flush Valve |
Recessed, self-closing, push-button, color coded and labeled, provides 45-55 lpm constant flow, while push-button is depressed; may not be used with O2 concentrator as O2 power source |
| CO2 Absorber System |
King Systems KAB-9 (re-fillable) or KAB-1 (pre-filled/disposable) |
| C02 Absorber Canister Capacity |
400 grams soda-lime |
| Directional Valves |
Built-in the CO2 absorber |
| CO2 Absorber Holding Bracket |
Plastic, secured with knob to main frame of machine |
| Bag-Ventilator Switch/PRV and Scavenger
Outlet Port |
Hand-operated selector switch and rotating knob for PRV and scavenging outlet |
| Bellows |
Latex free, upward inflating,
range from
0 to 1.6L |
| Bellows Pressure Relief |
Pre-set at 60 cmH20 |
| Common Gas Outlet |
Quick-connect, size indexed |
| Tubing Circuit |
King Systems F-360-61 or any standard anesthesia circle circuit |
| Gas Pressure Hoses |
DISS and thread indexed, female connectors at both ends |
| Gas Inlet Manifold |
DISS and thread indexed,male connectors with one-way valves |
| Oxygen and Air Supply
Gauges |
0-3000 PSI reading color
coded and
clearly labeled |
| Gas Inlet Manifold
Filters |
Located behind Manifold Air and O2 inlet male connectors |
Gas Inlet Pressure
Regulators
Main Supply Cylinder
Safety back up Cylinder |
DISS/thread indexed for O2 Pin-indexed, yoke mounted for "D" and "E" cylinders for O2 |
Oxygen Supply Alarms
|
|
| Main and Safety back-up |
0-3000 psi range, color coded and clearly labeled |
Alarm power source |
9-volt battery located in body of alarm box |
| Alarm on/off |
Labeled toggle switch located on body of alarm box |
| Air and O2 Flowmeters |
Calibrated and scaled 0-10 lpm, color coded, O2 flowmeter has a fluted control knob for easy identification by touch, alone |
| Oxygen Concentrator |
To power O2 flowmeter
only |
| Air Compressor |
May be used to power ventilator
and air flowmeter |
| Oxygen Analyzer/Monitor |
OM-25-ME (or equivalent) Galvanic cell sensor, life expectancy 2 years under normal conditions |
| Oxygen Analyzer Power
Source |
2 each AA batteries, life
expectancy approx. 3000 use hours |
| Auxillary
O2 Flow Selector |
Scaled
0-10 LPM in set increments, used for pre/post anesthesia |
| Vaporizer |
Penlon
SigmaDelta Series, bolt (cage) mounted, temperature
compensated, very low maintenance |
| Airway Pressure Gauge |
Dual scaled in cmH2O and mmHg, located on front panel of ventilator |
| Pressure Gauge Tubing |
May be attached to bag/vent switch arm or to a point within the patient breathing circuit (operators choice). |
| Mechanical
Ventilator |
Pneumatically powered, time cycled, volume constant, flow variable |
| Volume Range |
6 to 14 L |
| Insp. Flow Range |
10 to 40 L/min |
| Frequency |
breaths per minute 10 to 30 |
| Exp. Ratio |
Fixed 1:1.8 nominal |
| Ventilator
Gas Power Requirements |
40 to 70 psi, 50 psi optimal |
| Waste
Gas Scavenger |
Positive and negative relief valves, 1 L reservoir bag, vacuum control knob |
| Total
Machine Gas Leakage |
@ 30cm
H2O
@ 80cm H2O |
-0-ml/Min
-0-ml/min |
| Internal
System Compliance |
@ 20cm
H2O
@ 40cm H2O |
1.1ML/cm
H2O
1.3ML/cm H2O |
| Internal
System Resistance |
@ 1.0L/sec
gas flow
@ 0.5L/sec gas flow |
4.11cm
H2O
1.80cm H2O |
| APL
Valve Pressure Drop |
@ 3.0L/min
gas flow
@ 30L/min gas flow |
0.12cm
H2O
1.03cm H2O |
| Storage |
|
| Indoor |
+ 160
°F Allow unit to warm to normal |
Indoor
Outdoor |
-30
°F OR temp for 1hour |